SARS-CoV-2-Screen SARS-CoV-2 Real-time PCR Detection Kit

SARS-CoV-2-Screen kit is intended for the qualitative detection of SARS-CoV-2 coronаvirus RNA by RT-PCR in clinical samples (nasopharyngeal swab, oropharyngeal swab, sputum).

  1. Qualitative detection of SARS-CoV-2 coronаvirus RNA in clinical samples


  1. Target analytes: ORF1ab gene fragment of SARS-CoV-2 coronavirus, S gene fragment of SARS-CoV-2 coronavirus

Wielkość opakowania, ml REF Cena, €
IVD RK-SR-PCR-I 190
RUO RK-SR-PCR-R 150

Specyfikacje
Intended purpose SARS-CoV-2 real-time PCR detection
Number of tests 96 tests
Technology Reverse transcription polymerase chain reaction (RT-PCR) with real-time fluorescence detection; Qualitative assay
Equipment CFX-96 (Bio-Rad), QuantStudio 5 (Thermo Fisher Scientific), Rotor Gene 3000, 6000 (Corbett Research), Rotor-Gene Q (QIAGEN), DTlight 4S1, DTprime, DT-96 (DNA-Technology), or analogous
Compatible DNA/RNA isolation kits ExtraMag 1 kit, ExtraMag 2 kit, ExtraMag 3 kit гиперссылки
Clinical material Nasopharyngeal swab, oropharyngeal swab, sputum
Analytical specificity The potentially interfering substances do not interfere with the assay result*
Analytical sensitivity 5*102 copies of virus RNA (viral particles) per ml*
Diagnostic specificity 100% (95% CI: 86.3 ~ 100.0) *
Diagnostic sensitivity 100% (95% CI: 86.3 ~ 100.0) *
Reproducibility The coefficient of variation does not exceed 2%*
Storage conditions 2–8 ºC
Transportation conditions 2–8 ºC*
Shelf life 12 months
* See the user manual for the details